![]() |
![]() |
|
|
Ventricular Septal Defect (VSD) is the most challenging heart defect occlusion. Muscular VSDs can be occluded by most disk devices; however they are rare. Membranous VSDs are common; however they are very close to critical structures like the aortic valve and the tricuspid valve. Careful measurements of the VSD-right aortic cusp distance need to be made before a disk device is deployed. There is also procedural difficulty since the majority of the defects are crossed from the left side, an arterio-venous connection needs to be made and they are corrected from the venous side. The femoral vein is used for the outflow defects (membranous) and the right jugular vein for the mid-muscular defects. The buttoned device and its different modifications has been used for the occlusion of congenital muscular VSDs, post-infarction VSDs and several membranous VSDs after the first year of life. The results have been good. 9-11 F introduction is required The transcatheter patch has the potential to revolutionize VSD occlusion, since most membranous VSDs can be repaired. The patch requires a minimal rim and therefore can occlude a defect without interference with the aortic valve. The VSD patch requires 9-10F introduction.
|
|
TRANSCATHETER CORRECTION OF VENTRICULAR SEPTAL DEFECTS BY THE “SIDERIS” DEVICES E.B.
Sideris M.D. Athenian
Institute of Pediatric Cardiology and Custom Medical Devices Introduction Transcatheter ventricular septal defect occlusion is by far the most challenging heart defect occlusion. The reasons are both anatomic and technical. Most ventricular septal defects are peri-membranous, in close proximity to critical structures like the aortic valve and the tricuspid valve( 1); most of these defects have deficient rim for disk device placement . Muscular VSDs are the ones traditionally considered as good candidates for transcatheter occlusion (2 ); they usually have enough supportive tissue(rim) for disk device placement ; they are also considered high risk for surgery, since they cannot always be approached from the tricuspid valve and require ventriculotomy ( 3 ). The technical limitations are significant and include: a. Difficulty crossing the defect: The easiest approach to cross a VSD is the transarterial approach. The size and the location of the defect are the most important factors; smaller than 3 mm defects and supracristal defects are the most difficult. Large mid-muscular defects can be frequently crossed directly transvenously, from the right jugular vein.
soft wire needs to be advanced into the pulmonary artery and
snared to the most
appropriate vein (femoral for outflow defects, right jugular
for mid-muscular
ones).
VSDs; they certainly have an advantage with multiple defects,
because of
significant overlapping; in contrast disk devices are
applicable in only a small
percentage of membranous VSDs ( 5
), because of the
close proximity to critical
structures; careful measurements are required to select the
appropriate disk
device size. The development of wireless devices (transcatheter
patch and
detachable balloon) which require minimal rim and cannot
interfere with critical
structures (6,7) has the potential to revolutionize the
transcatheter treatment of
peri-membranous VSDs.
long delivery sheaths. A straight rather an angulated route, is preferable to avoid
kinking of the sheath. Apical muscular VSDs are rather
notorious in this aspect.
Over the wire placement of the device and kink-resistant
sheaths can be helpful in
difficult VSD locations. The Devices Since 1994, the following devices have been used by us for the occlusion of VSDs:
Buttoned
Device: This is the same device used for the occlusion of
atrial septal defects ( 8
). The current device is a 4th generation device
without centering since VSDs are relatively small in size. The device
has been described extensively. In summary it consists off an occluder
with the button loop, a counter-occluder and a loading or release
wire. It can be placed directly through a long sheath or over a wire.
The over the wire placement possibility makes the device quite
attractive since chances to loose the arterio-venous connection in
case of device mis-placement or retrieval are minimized. The occluder
is placed from the left ventricular side, the button-loop adjusts for
the septal thickness and the counter-occluder is buttoned with the
occluder from the right ventricle. The
Self-Adjustable Device: The
device consists off two disks (proximal and distal) connected by a
latex thread (9). It can therefore self-adjust according to the septal
thickness (Fig 1). The disks are made by polyurethane foam and
stainless steel skeleton
wire. There are two wire and one wire disks
in different combinations as follows (Fig 2): a.
regular
self-adjustable device, with a two wire distal disk and a one wire
proximal disk. b.
inverted
self-adjustable device, with a single wire distal disk and a two wire
proximal disk . This
is the device most commonly used for transarterial VSD
occlusion. c.
two
disk device, both the proximal and the distal disks are made with two
wires. d.
One
wire device, both the proximal and the distal disks are made with a
single wire. Wireless
Devices : Two
types of wireless occluders have been used in the occlusion of VSDs.
Detachable balloons and Transcatheter patches (6,7 ). These devices
have been described extensively in a different chapter. In
Summary the detachable balloon device consists off a detachable
balloon occluder, connected with a floppy disk. The detachable balloon
occludes the defect from the left ventricle and the floppy disk(s)
supports the balloon from the right ventricle. The transcatheter patch is made by polyurethane foam in the form of a sleeve over a balloon. It is retrievable, retractable and radiopaque. It is delivered and supported by a double balloon for 48 hours. It is totally wireless. Both
detachable balloons and transcatheter patches require minimal rim and
are always centered; however the detachable balloon method is an
outpatient procedure where the transcatheter patch application
requires 48 hours hospitalization.
Patient Selection Indication
for VSD occlusion is the persistence of
a significant shunt with
or without evidence of pulmonary hypertension.
Exceptions for occlusion include irreversible pulmonary
hypertension with right to left shunt (PVR> 10 units) and anatomy
preventing the application of the device (lack of sufficient rim). Calculation
of adequate rim and examples are given in the included diagram (Fig3).
The same rim requirements exist
for both disk devices used. The wireless devices require a minimal
rim. Balloon test occlusion should be performed prior to a wireless
device application (transcatheter patch). Providing that it results in
full occlusion without impairment of critical structures, patch
deployment should follow. Ventricular
septal defects of all types can be occluded provided there is adequate rim for the deployment of the device.
Baseline history, physical exam,
EKG , echocardiogram and chest x-ray need to be obtained for
screening purposes. Exclusion
Criteria: 1.
Defects
with severe pulmonary hypertension associated with significant R-L
shunts. (PVR>10 units) 2.
Defects
without adequate rim for device deployment 3.
Patients
too small for a possible retrieval of the device into a 11 F sheath
(10 Kg) for disk devices or
9-10 F for the transcatheter patch.
Surgical retrieval
should be performed in
patients where the appropriate sheath for the retrieval
cannot be introduced Device Selection The
disk devices selected should
have at least twice the size of the diameter of the VSD (similarly to
the ASD selection). Caution
should be applied in the selection of the proper device to avoid
encroachment to the aortic valve leaflet. A long axial
angiographic view should be obtained and the distance between the
center of the defect and the right aortic cusp should be measured (DA)
as well as the diameter of the defect (Fig 4,5 ). The diameter of the
required device should be less than double that of the DA distance and
be at least 2mm away from the right aortic cusp.
The minimal sub-aortic rim (DA) for the smaller available
device (15mm) is calculated therefore at 9.5mm. The
transcatheter balloon/patch
diameter should be selected as 2mm larger than the test occluding
diameter (Fig 6,7). The Method The
procedure is performed in the cardiac catheterization laboratory. The
patient is pre-medicated with the
Toronto cocktail .
Sedation is supplemented by Ketamine infusion. General
anesthesia is reserved for the cases requiring transesophageal
echocardiography or simple sedation is not considered adequate. The
procedure should be performed using sterile technique. After obtaining
routine pressure and oxygen saturation data to confirm the diagnosis
of VSD, angiography is performed , including long axial left
ventricular angiogram . The
location of the defect is
noted and the size is digitally measured (Fig 2). In
addition to the size, the distance from the center of the defect to
the right aortic cusp is measured and the appropriate disk device is selected. If the distance is too small for a disk device the
use of a wireless device is entertained.
Heparinization is started (100 units/Kg). The
next step includes crossing the VSD by a catheter. Crossing from the
right side is preferable, but not always possible; therefore in the
majority of cases the defect is crossed from the left ventricle,
utilizing a right Judkins or a modified Amplanz catheter.
Transvenous
approach : A
regular 025” 260 cm wire is utilized and it is advanced in the
pulmonary artery from
where it is snared to the femoral vein (membranous defects) or the
jugular vein (muscular defects).
Attention is paid to avoid bleeding from the arterial and the
venous sites. A
long sheath in a size appropriate to the device used is advanced over
the wire and is positioned in the ascending aorta.
Buttoned
Device Placement: The
occluder is introduced over the wire into the sheath and it is
advanced to the aortic valve
level, where releasing is started ; however the sheath is slowly
pulled back until the entire occluder is released into the left
ventricle . The occluder
is manipulated away from the aortic valve and is aligned with the
ventricular septum. Echocardiography confirms the good position and
the occlusion of the
defect. The counter-occluder is buttoned
as usual and a repeat left ventricular angiogram is performed
(Fig 5). The
wire is withdrawn through the device with the tip of the sheath
placed against the device. The device is subsequently released. Transvenous
approach using the transcatheter patch: A
long sheath in a size appropriate for the transcatheter patch used is
advanced over the wire in the ascending aorta. The wire is taken out.
Both the distal and the proximal balloons are entering the
ascending aorta. The distal balloon/ patch is inflated
and it is pulled until it obstructs the defect; fluoroscopy,
echocardiography and pressure recording are guiding the procedure.
After full occlusion is confirmed the proximal balloon in
inflated. Subsequently,
the balloon catheter and the long sheath are immobilized on the skin.
A left ventricular angiogram and an aortogram confirm the result (Fig
7,8). Transarterial
Approach Using the Self –Adjustable Device. If
the size and the location of the
ventricular septal defect are
considered appropriate for disk device occlusion , a 0.025” stiff exchange wire is placed at the apex of the
right ventricle. An
appropriate size sheath is placed over the wire in the right
ventricle. An inverted self adjustable device is loaded and it is
advanced until the distal (single wire disk) is released in the right
ventricle; subsequently the loading wire of the device is pulled until
the distal disk becomes perpendicular to the tip of the sheath. The
whole complex is pulled back until it stops at the ventricular septal
level. The sheath is pulled back in the left ventricle and the
proximal disk is released, occluding the VSD. The proximal disk could
be manipulated away from the aortic valve; the distance between
proximal and distal disk is self-adjusted according to septal
thickness (Fig1). The device is subsequently released in a similar fashion to
the buttoned device. Oximetry
will be reserved to patients with residual shunts , but it should be
done carefully to avoid disturbing the device.
(Qp:Qs=1 no shunt, Qp:Qs 1-1.2=trivial shunt, 1.3-1.5=small
residual shunt, Qp:Qs>1.5 large shunt.) Antibiotic
prophylaxis should be started (preferably Cephalosporins with the
first dose given IV in the cath lab and the subsequent ones orally six
and twelve hours later for the disk devices.
Antibiotics are continued for 72 hours for the transcatheter
patch. The patient should be discharged home within 24 hrs from the
release of the device. Echocardiogram,
chest x-rays and EKG will be obtained prior to discharge.
The patient should
be discharged home with
daily medication of Aspirin
(300mg) for six weeks and endocarditis prophylaxis for six months for
full occlusions; in case of residual shunts endocarditis prophylaxis
will continue indefinitely. Results Fifty
five ventricular septal defects have been occluded by the
“Sideris” devices since 1994. The majority of the defects (36)
were occluded by the buttoned device, transvenously . The
transarterial approach with the self-adjustable device was used in 14
cases. Wireless devices were used in 5 cases. Most
occluded defects were peri-membranous (45); five congenital muscular
and five post-infarction VSDs were occluded as well. One of the
congenital muscular VSDs was swiss-cheese type. The
size of the VSDs varied from 3-12mm (m=5) for the congenital and
12-25mm for the post-infarction. Patient age varied from 1 to 45 years
(m=7) for the congenital VSDs and 65-75 years (m=70) for the
post-infarction. The average procedure time was 2.5 hours for
transvenous procedures and 30
minutes for transarterial procedures utilizing the self-adjustable
device. However transarterial procedure could not be performed in 3
attempted cases which were performed transvenously instead.
Wireless occlusions were performed in membranous VSDs
inappropriate for disk device occlusion (three detachable balloons,
two transcatheter patches).
There
was immediate effective occlusion in all cases. Three early cases
occluded transvenously with the buttoned device developed mild aortic
insufficiency. In 2 the devices were extracted surgically in 48 hours.
There was a transient 3rd degree A-V block in one case and
transient PVCs in most cases during implantation. Despite
early symptomatic improvement only 2/5 cases after myocardial
infarction are long term survivors.
No complications were seen with the self-adjustable device. One
detachable balloon leaked in 24 hours and had to be extracted
surgically. ComplicationsSerious
complications requiring re-intervention were the two early buttoned
device cases which developed aortic insufficiency and the case of
the detachable balloon early leakage, which required
extraction; however we are listing a series of potential
complications ; this list should be useful in obtaining an informed
consent from the patients. Complications
as seen in the ASD trials and the international VSD trials are defined
as follows: 1:
unbuttoning and embolization of the occluder on the left or on the
right side of the heart (self explanatory) 2:
cardiac perforation: perforation of a cardiac chamber by a catheter, a
wire or a device component during the procedure 3:
air embolism : introduction of air in the cardiac cavities from the
introducing sheath, with findings of
transient ST elevation, bradycardia and transient drop of the
systemic pressure; this a self limited phenomenon within 5-10 minutes,
seeing rarely during ASD occlusion. Such an episode is unlikely for
VSD occlusion, since high pressure bleeding prevents air entrapment 4:
excessive bleeding: small
bleeding from the entry sites is not uncommon, especially if an
arterio-venous connection is
established ; the investigator should be ready
for the appropriate measures (sealing with pressure, new valve
sheaths etc) . Blood transfusions might be
needed. 5:
transient arrhythmia : PVCs are common during catheter, wire or device
introduction in the ventricles; they are usually self-limited.
Transient heart block has been seen as well. 6:
mitral insufficiency : it has been seen in ASD occlusion; highly
unlikely with VSD occlusion 7:
tricuspid insufficiency: trivial tricuspid insufficiency has been seen
without problems; in case of significant tricuspid insufficiency the
device should be extracted. 8:
aortic insufficiency : it is the commonest problem after membranous
VSD occlusion, requiring extraction; it can be prevented by proper
patient/device selection. 9:
peripheral vascular problems .
Discussion The concurrent use of three device types, illustrates the inherent difficulties of VSD occlusion. The muscular VSDs were unquestionably the easiest to close by disk devices. Both the buttoned device and the self-adjustable device were used. The advantage of the significant overlapping of the disk devices could be best illustrated in the occlusion of a “swiss-cheese” type VSD by the buttoned device. The
size of the self-adjustable device is restricted to 25mm; therefore
the use of the buttoned device was preferable for the large
post-infarction muscular VSDs. Apical muscular VSDs are rather
difficult for buttoned device delivery, with the long sheath kinking
in two occasions. Over
the wire placement and kink-resistant sheaths was the solution of this
problem. Despite the successful post-infarction VSD occlusion in 5
cases and their symptomatic improvement only two of them are long term
survivors. The necessity
of a stressful and high risk procedure with questionable long-term
outcome is raised. The
main procedural difficulty for transvenous VSD occlusion is crossing
the VSD and establishing an arterio-venous connection. The possibility
of over the wire
placement with the
buttoned device makes the procedure more secure. On
the other hand if the procedure can be done trans-arterially using the
self-adjustable device, it
is much faster; the average procedural time is only 30 minutes in
comparison to 2.5 hours for the transvenous procedure. The
majority of ventricular septal defects are peri-membranous; this is
the reason we focused our research effort in this category and the
majority of our occlusions were related to peri-membranous defects (
10 ). The first and the
most commonly used
device for peri-membranous VSD occlusion was the buttoned
device. The use of oversized devices early in the clinical trial, was
responsible for three instances of aortic insufficiency; two of them
required surgical
extraction of the device. Careful
measurements of the defect to right aortic cusp distance and selection
of smaller devices positioned at
least 2mm away from the aortic cusp was the solution of the problem;
indeed no more cases of aortic insufficiency were noticed. The
presence of an aneurysm of the membranous septum was helpful in
several cases since the occluder could be pulled away from the aortic
valve in the aneurysm. The
use of disk devices in the peri-membranous VSD occlusion is limited;
and the majority of such VSDs are sent to surgery. The evolution of
the wire-less devices has the potential to change this practice. The 5
cases performed by us are in favor of
this projection. All of them were inappropriate for disk device
occlusion and all of them were surgical candidates. Both
wireless devices used, are working on a balloon occluding principle. A
balloon is a 3-dimensional structure requiring a minimal rim for
support; a sub-aortic rim as small as 1 mm in enough for the
occlusion. One
detachable balloon device used leaked prematurely; it embolized in the
pulmonary artery and had to be extracted surgically. The
transcatheter patch offers the efficacy of the detachable balloons but
unparallel safety with double cardiac and extracardiac immobilization
and retractability in the introducing sheath. Despite the need for 48
hour balloon support, this wireless method has the potential to become
the procedure of choice for the majority of peri-membranous VSDs. Current StatusThe
use of devices for the transcatheter occlusion of VSDs is
investigational. All such devices are used under protocols and
regulations acceptable in each center. The
FDA has approved clinical trials with the buttoned device in the
occlusion of ventricular septal defects, in USA. Both peri-membranous
and muscular defects are occluded under the approved protocols. An
Investigational Device Exception for the transcatheter patch use for
the occlusion of heart defects has been submitted to the FDA. ConclusionsTranscatheter VSD occlusion using the different “Sideris” devices has been applied successfully since 1994. Three types of devices have been used to optimize the results. The
buttoned device is the commonest one used, over an arterio-venous
wire, for transvenous occlusion. The
self-adjustable device has
the easiest application since it can be applied transarterially. Wireless
devices and specially the transcatheter patch are the most promising
methods for transvenous peri-membranous
VSD occlusion. References 1. Graham TP and
Gutgesell HP. Ventricular septal defects. In Moss and Adams-
Heart Disease in Infants, Children, and Adolescents, Fifth Edition ,
1995: 724-746 2. Lock JE, Block PC, McCay RG et al.
Transcatheter closure of ventricular septal defects. Circulation
1988:78:361-8
3.Bridges ND, Perry SB, Keane JF et al. Preoperative
transcatheter closure of
congenital muscular ventricular septal defects. N
Engl J Med 1991;324:1312-7
4.Thanopoulos BD, Tsaousis GS, Konstantopoulou Gl et al.
Transcatheter closure of
muscular ventricular septal defects with the amplantzer
ventricular septal defect
occluder: initial clinical application in children. J Am. Coll
Cardiol 1999;33:1395-9
5.Kalra GS, Verma PK, Dhall A et al. Transcatheter device
closure of ventricular
septal defects: immediate results and intermediate term
follow-up. Am Heart J
1999; 138:339-44
6. Sideris EB, Kaneva A, Sideris SE, Moulopoulos
SD. Transcatheter atrial septal
defect occlusion in piglets, by balloon detachable devices.
Cath Cardiov Interv
2000; in press
7. Sideris E, Toumanides S, Alekyan B et al. Transcatheter
patch correction of atrial
septal defects: Experimental validation and early clinical
experience.
Circulation 2000; 102:Suppl.II-588
8. Sideris EB, Sideris SE, Thanopoulos BD et al. Transvenous
atrial septal defect
occlusion by the buttoned device. Am. J Cardiol 1990;66:1524-6
9. Sideris EB, Zheng J, Wang Y, Haddad G. Transarterial
occlusion of membranous
ventricular septal defects by a self-adjustable device.
Circulation 1998;98:I-755
10. Sideris EB, Walsh KP, Haddad JL, Chen C-R, Ren SG, Kulkarni
H. Occlusion of
congenital ventricular septal defects by the buttoned device.
Heart 1997; 77:276-
279
Figure LegentsFigure 1: The Self-Adjustable DeviceFigure
2: The different types of self-adjustable device: 1. Regular 2.
Inverted 3. Two
Disk, 4. Single disk Figure
3: Angiographic measurements: 1. VSD diameter, 2. VSD-Right Aortic
Cusp
Dimension Figure
4: Example of Angiographic
Measurements and Device Selection for
Perimembranous VSD: D1= VSD diameter, 5.1 mm, D2= VSD- right
aortic
cusp dimension 13mm; any device 25mm or smaller could be used.
Figure 5: Occlusion of a
5mm perimembranous defect (same case of Fig.4) by a 20 mm
buttoned device with full occlusion. Figure
6: Large perimembranous
VSD (12 mm) in a 4 year old child, very close to the
aortic valve (3mm) Figure
7: Transcatheter Patch
Occlusion supported by a double balloon 14mm in
diameter; same case as Figure 6. Figure
8: Aortogram shows no interference to the aortic valve.
|
|
|
Copyright 2001-2005. CustomMedicalDevices.com